WEBVTT

00:00:00.400 --> 00:00:02.210
Welcome to Ask GAEA.

00:00:03.245 --> 00:00:06.150
This guide shows how to enter falling-head permeability

00:00:06.165 --> 00:00:09.390
readings in GaeaSynergy and check the reported

00:00:09.405 --> 00:00:10.970
hydraulic conductivity.

00:00:12.325 --> 00:00:16.090
From the project Dashboard, open the sample and confirm

00:00:16.105 --> 00:00:17.850
the ID against the worksheet.

00:00:18.685 --> 00:00:21.450
Go to Geotechnical Tests and open Falling Head

00:00:21.465 --> 00:00:24.570
Permeability, or add a new test with the matching

00:00:24.585 --> 00:00:30.970
template. Open Edit, Test Data to review the setup.

00:00:34.205 --> 00:00:37.750
On Falling Head Information, set the units for time,

00:00:37.865 --> 00:00:40.730
length, temperature, mass and conductivity.

00:00:43.965 --> 00:00:46.890
Enter the specimen dry mass, height and diameter.

00:00:47.905 --> 00:00:50.850
Record the standpipe diameter and initial water height.

00:00:53.625 --> 00:00:56.390
Verify the specific gravity and confirm the reference

00:00:56.405 --> 00:00:58.690
used to measure every water height.

00:00:59.325 --> 00:01:03.410
On Falling Head Data, enter increasing elapsed times,

00:01:04.185 --> 00:01:05.810
measured water heights and temperatures.

00:01:09.705 --> 00:01:12.130
Use the initial height field for time zero.

00:01:12.825 --> 00:01:14.870
Leave the last cell to recalculate.

00:01:17.465 --> 00:01:20.110
Compare one row with the worksheet using consistent

00:01:20.125 --> 00:01:21.670
length and time units.

00:01:22.665 --> 00:01:25.890
Keep the average, temperature-corrected average and

00:01:25.905 --> 00:01:27.510
regression fit distinct.

00:01:31.745 --> 00:01:33.650
Click OK to view the report.

00:01:34.625 --> 00:01:37.530
In Height of Water versus Time, find the measured

00:01:37.545 --> 00:01:39.550
points and check them against the worksheet.

00:01:39.565 --> 00:01:45.970
In Hydraulic Conductivity versus Time, match a plotted

00:01:45.985 --> 00:01:48.270
point and the average to the results.

00:01:49.005 --> 00:01:51.930
Save and reopen the test to confirm the readings,

00:01:51.985 --> 00:01:53.930
calculations and plots persisted.

00:01:57.265 --> 00:01:59.930
You now have a checked falling-head permeability report

00:01:59.945 --> 00:02:01.430
with its original readings.

00:02:02.865 --> 00:02:06.746
For more walkthroughs, open Ask GAEA Guides in GaeaSynergy.
